Description
Flame Photometer LFP-A41 can consume 6ml of sample per minute and works in a temperatu range of 10 to 35 °C. It measures Potassium (K), sodium (Na), lithium (U), calcium (Ca), and barium (Ba) concentrations. Utilizes silicon photovoltaic cells as photoelectric conversion elements. It offers a direct reading function for measuring the concentration of potassium and sodium. With a flameout protection device, our flame photometer is safe to operate.
- Specifications :
Testing Elements K, Na, Li, Ca, Ba
Sample Volume < 6mL/min
Temperature Range 10 to 35 ℃
Relative Humidity ≤ 85%
K Measurement Range 0 to 100ug/ml
Na Measurement Range 0 to 160ug/ml
Li Measurement Range 0 to 100ug/ml
Ca Measurement Range 0 to 1000ug/ml
Ba Measurement Range 0 to 3000ug/ml
Stability ≤15%
Repeatability ≤ 3%
≤0.005mmol/L (0.0100~0.0800)
mmol/L K Linearity Error
Na Linearity Error ≤0.03mmol/L (0.0500~0.400) mmol/L
Li Linearity Error ≤0.021mmol/L
Ca Linearity Error ≤0.075mmol/L
Ba Linearity Error ≤0.066mmol/L
K Detection Limit ≤ 0.004mmol/L
Na Detection Limit ≤ 0.008mmol/L
Li Detection Limit ≤0.015mmol/L
Ca Detection Limit ≤0.050mmol/L
Ba Detection Limit ≤0.044mmol/L
Peak Wavelength Error: ≤ 7nm
Half width≤15nm
Background Transmittance≤ 0.5%
K, Li, Ca, Ba Filter
Peak Wavelength Errors: 5nm
Half-Width: ≤ 15nm
Background Transmittance: ≤ 0.5%
Na Filter
Spectroscopic Methods Interference Filter
Response Time <8s
Display LCD screen display
Reviews