Flame Photometer LFP-A41

Feature

  • Use liquefied petroleum gas as fuel
  • Operated using a menu-driven keyboard
  • Automatic calculation function
  • Option for pre-select the flame size
  • Data can be printed directly
  • Computer online data processing system included
Description

Description

Flame Photometer LFP-A41 can consume 6ml of sample per minute and works in a temperatu range of 10 to 35 °C. It measures Potassium (K), sodium (Na), lithium (U), calcium (Ca), and barium (Ba) concentrations. Utilizes silicon photovoltaic cells as photoelectric conversion elements. It offers a direct reading function for measuring the concentration of potassium and sodium. With a flameout protection device, our flame photometer is safe to operate.

  • Specifications :
    Testing Elements K, Na, Li, Ca, Ba
    Sample Volume < 6mL/min
    Temperature Range 10 to 35 ℃
    Relative Humidity ≤ 85%
    K Measurement Range 0 to 100ug/ml
    Na Measurement Range 0 to 160ug/ml
    Li Measurement Range 0 to 100ug/ml
    Ca Measurement Range 0 to 1000ug/ml
    Ba Measurement Range 0 to 3000ug/ml
    Stability ≤15%
    Repeatability ≤ 3%
    ≤0.005mmol/L (0.0100~0.0800)
    mmol/L K Linearity Error
    Na Linearity Error ≤0.03mmol/L (0.0500~0.400) mmol/L
    Li Linearity Error ≤0.021mmol/L
    Ca Linearity Error ≤0.075mmol/L
    Ba Linearity Error ≤0.066mmol/L
    K Detection Limit ≤ 0.004mmol/L
    Na Detection Limit ≤ 0.008mmol/L
    Li Detection Limit ≤0.015mmol/L
    Ca Detection Limit ≤0.050mmol/L
    Ba Detection Limit ≤0.044mmol/L
    Peak Wavelength Error: ≤ 7nm
    Half width≤15nm
    Background Transmittance≤ 0.5%
    K, Li, Ca, Ba Filter
    Peak Wavelength Errors: 5nm
    Half-Width: ≤ 15nm
    Background Transmittance: ≤ 0.5%
    Na Filter
    Spectroscopic Methods Interference Filter
    Response Time <8s
    Display LCD screen display
Reviews (0)

Reviews

There are no reviews yet.

Be the first to review “Flame Photometer LFP-A41”